Clinical Research Directory
Browse clinical research sites, groups, and studies.
Fracture Monitor - Femur
Sponsor: AO Innovation Translation Center
Summary
Thirty-seven patients diagnosed with femoral fracture to be treated with a compatible plate will be enrolled for this multicenter prospective interventional study. The Fracture Monitor T1 will be attached to the locking plate during osteosynthesis. Prior to discharge, it will be connected to an app installed either on the patients' smartphone or on the smartphone provided by the sponsor. The device will continuously collect and process data, which can automatically be downloaded via Bluetooth through the smartphone app and transferred in encrypted form to a dedicated cloud server. The smartphones will be equipped with a basic app version with rudimentary user interface that has no data visualization function. During the rehabilitation phase, data collected by the device will not be accessible to the surgeon nor the patient to ensure that no therapeutic decision will be made based on the acquired data. Every patient is treated as per standard of care. The primary objective of the study is to collect safety information of the Fracture Monitor T1. The secondary objectives are to collect information on device performance, device handling, and usability. All adverse events (AEs) and serious AEs (SAEs) will be recorded during the study. These will be reviewed and evaluated for their potential relationship to the device. All patients will be followed up for 6 months after the surgery according to the local standard of care. Given that the removal of the Fracture Monitor T1 may occur beyond 6 months, a safety follow-up will take place at the time point when the Fracture Monitor T1 is removed as per the instructions for use if it has not been removed within the 6 months. The results will be compiled into a clinical evaluation report to be submitted as part of the technical documentation for the conformity assessment procedure.
Official title: Prospective First in Human Clinical Investigation to Evaluate the Safety of the Fracture Monitor T1 in Patients with Femur Fractures Treated with a Locking Compression Plate
Key Details
Gender
All
Age Range
18 Years - Any
Study Type
INTERVENTIONAL
Enrollment
37
Start Date
2023-10-03
Completion Date
2026-07
Last Updated
2025-02-03
Healthy Volunteers
No
Conditions
Interventions
Fracture Monitor (implantable device class III)
Femoral fracture fixation (bridge plating aiming at secondary bone healing) using a locking plate with an attached investigational Fracture Monitor T1.
Locations (4)
Universitätsklinikum des Saarlandes
Homburg, Germany
Universitätsklinikum Münster
Münster, Germany
Berufsgenossenschaftliche Unfallklinik Tübingen
Tübingen, Germany
Universitätsklinikum Ulm
Ulm, Germany