Clinical Accuracy Validation of the Moni-Patch Temperature Monitoring Patch System

Sponsor: Murata Vios, Inc.

View on ClinicalTrials.gov

Summary

Moni-Patch Temperature Monitoring Patch System, a non-invasive, continuous temperature monitoring device designed to estimate core body temperature by applying a Sensor to the neck. This clinical accuracy validation study aims to verify the accuracy and repeatability of the Moni-Patch compared with an FDA-cleared continuous tympanic temperature monitoring device used as the clinical reference. Main objectives of this study are; To verify that the Moni-Patch and the reference device(tympanic) demonstrate acceptable agreement, as defined by Bland-Altman Analyses (bias within ±0.4°C and limits of agreement between -1.0°C and 1.0°C). To verify the consistency and repeatability of temperature measurements obtained from the Moni-Patch during continuous monitoring.

Official title: Moni-Patch Temperature Monitoring Patch System Clinical Accuracy Validation

Key Details

Gender

All

Age Range

7 Years - Any

Study Type

INTERVENTIONAL

Enrollment

105

Start Date

2026-01-21

Completion Date

2026-03-31

Last Updated

2026-02-09

Healthy Volunteers

Yes

Interventions

DEVICE

Moni-Patch Temperature Monitoring Patch System

Wireless neck-applied Sensor paired with a Receiver via Bluetooth; heat flux and thermistor-based estimation of core temperature; sampling every 30 seconds.