Clinical Accuracy Validation of the Moni-Patch Temperature Monitoring Patch System
Sponsor: Murata Vios, Inc.
Summary
Moni-Patch Temperature Monitoring Patch System, a non-invasive, continuous temperature monitoring device designed to estimate core body temperature by applying a Sensor to the neck. This clinical accuracy validation study aims to verify the accuracy and repeatability of the Moni-Patch compared with an FDA-cleared continuous tympanic temperature monitoring device used as the clinical reference. Main objectives of this study are; To verify that the Moni-Patch and the reference device(tympanic) demonstrate acceptable agreement, as defined by Bland-Altman Analyses (bias within ±0.4°C and limits of agreement between -1.0°C and 1.0°C). To verify the consistency and repeatability of temperature measurements obtained from the Moni-Patch during continuous monitoring.
Official title: Moni-Patch Temperature Monitoring Patch System Clinical Accuracy Validation
Key Details
Gender
All
Age Range
7 Years - Any
Study Type
INTERVENTIONAL
Enrollment
105
Start Date
2026-01-21
Completion Date
2026-03-31
Last Updated
2026-02-09
Healthy Volunteers
Yes
Conditions
Interventions
Moni-Patch Temperature Monitoring Patch System
Wireless neck-applied Sensor paired with a Receiver via Bluetooth; heat flux and thermistor-based estimation of core temperature; sampling every 30 seconds.
Locations (1)
Murata Vios
Woodbury, Minnesota, United States