Clinical Research Directory
Browse clinical research sites, groups, and studies.
Controlled Oxygen Desaturation Study of the Neonatal Wireless Transmission System
Sponsor: Kathryn Beardsall
Summary
This study aims to test the oxygen saturation component of the newly developed NeWTS monitoring system and compare readings from NeWTS to a traditional wired clinical monitor. It is important the investigators do this study in healthy adults as part of the development of this system, to make sure the oxygen saturation readings are accurate and before NeWTS is used in Neonatal Intensive Care (NICU).
Official title: Controlled Oxygen Desaturation Study of the Neonatal Wireless Transmission System (NeWTS): NeWTS-DS
Key Details
Gender
All
Age Range
18 Years - 55 Years
Study Type
INTERVENTIONAL
Enrollment
24
Start Date
2027-01-11
Completion Date
2027-08-31
Last Updated
2026-09-10
Healthy Volunteers
Yes
Conditions
Interventions
Neonatal Wireless Transmission System
The primary objective for this clinical investigation is to validate the SpO2 performance of the NeWTS in comparison to "gold-standard" measurements of blood SaO2 by a calibrated and QC'ed ABL90 Flex plus blood gas analyser.
Locations (1)
Clinical Research Facility
Birmingham, England, United Kingdom