Operator-Assisted Oscilla Audiometry as a Basis for Hearing Aid Fitting (The Oscilla Validation Study)

Sponsor: Aalborg University Hospital

View on ClinicalTrials.gov

Summary

The purpose of this study is to find out whether hearing tests performed with the Oscilla A60 audiometer, where the patient largely tests themselves with minimal staff assistance, are accurate enough to be used for diagnosing hearing loss and as the sole basis for fitting hearing aids. Adults referred for their first hearing aids at Aalborg University Hospital will have their hearing tested both with the Oscilla audiometer and by a trained audiology assistant at the same visit. Participants are then randomly assigned to receive hearing aids fitted based on either the Oscilla hearing test or the audiology assistant's hearing test. The main questions are: (1) Do the Oscilla results agree with the audiology assistant's results for both air-conduction and bone-conduction hearing thresholds? (2) Do participants fitted based on the Oscilla test report a hearing benefit that is not worse than that of participants fitted based on the standard test, two months after fitting?

Official title: Validation of Operator-Assisted Oscilla Audiometry as a Diagnostic Tool and Basis for Hearing Aid Fitting in Adults With Uncomplicated Hearing Loss: A Blinded Non-Inferiority Randomised Controlled Trial

Key Details

Gender

All

Age Range

18 Years - Any

Study Type

INTERVENTIONAL

Enrollment

268

Start Date

2026-12-01

Completion Date

2027-10-31

Last Updated

2026-11-19

Healthy Volunteers

No

Interventions

DIAGNOSTIC_TEST

Paired audiometry (Oscilla A60 and manual audiometry)

At Visit 1, all participants undergo both operator-assisted Oscilla A60 audiometry (AudioConsole v4.6.0; AC and BC thresholds; operator places the BC headband only) in a quiet room (\<50 dB SPL) and manual pure-tone audiometry (AC and BC, ISO 8253-1) by a qualified audiology assistant in a sound-attenuating booth, in randomised order.

DEVICE

Oscilla-based hearing aid fitting

Bilateral BTE/RIC hearing aid fitting to NAL-NL2 targets using the Oscilla A60 audiogram as the sole audiometric basis, verified with auto-REM (within ±5 dB of target). Other names: Oscilla A60 Diagnostic Audiometer.

DEVICE

Manual audiometry-based hearing aid fitting

Bilateral BTE/RIC hearing aid fitting to NAL-NL2 targets using the audiology assistant audiogram, verified with auto-REM (within ±5 dB of target). Standard clinical practice.